36
EnteraLite
®
Infinity
™
Operator’s Manual
© Copyright 2004 ZEVEX Incorporated
1. Pump Components
5
Disposable Set
The only disposable sets approved for use with EnteraLite
®
Infinity
™
pump are:
EnteraLite
®
Infinity
™
500 ml Bag Set, Order Number INF0500
EnteraLite
®
Infinity
™
1200 ml Bag Set, Order Number INF1200
EnteraLite
®
Infinity
™
Spike Set, Order Number INF0010
Figure 1-10 EnteraLite
®
Infinity
™
Spike Set
Figure 1-9 EnteraLite
®
Infinity
™
500ml or 1200ml Bag Set
5
7
6
8
9
2
3
4
1. “ “ Symbol
2. Cassette
3. Downstream Tubing
4. Upstream Tubing
5. 500ml or 1200ml Bag
6. Barbed Enteral Adapter
7. Adapter Protective Cover
8. Spike Protective Cover
9. Spike
1
8. ADDITIONAL TECHNICAL INFORMATION
8. Additional Technical Information
Operational Conditions
EnteraLite
®
Infinity
™
is designed in accordance with EN 60601-1-2, EN
60601-1-4, and RTCA DO-160D standards for electromagnetic emissions and
immunity. Normal operation should not interfere with nor be affected by
other electronic equipment. EnteraLite
®
Infinity
™
can safely be operated on
commercial aircraft.
EnteraLite® Infinity™ Disposable Set Displacement
EnteraLite
®
Infinity
™
Disposable Set displaces approximately 15ml of fluid
when primed. Approximately 3.5 ml of air is drawn into the tubing before
the enteral pump detects that bag is empty. Therefore, approximately 11.5
ml of fluid remains in the disposable set when the pump stops feeding.
(These values vary by approximately 0.5 ml, depending on whether battery
is low or fully charged).
Routine Maintenance
There are no user serviceable parts or routine calibration or adjustment pro-
cedures required for EnteraLite
®
Infinity
™
. All service requests should be
referred to ZEVEX technical service.
25356_pages.qxd 2/2/2005 2:49 PM Page 11