2. Directions for Use
20
EnteraLite
®
Infinity
™
Operator’s Manual
Following is a list of all alarms, display messages, and indications that are used
by the EnteraLite
®
Infinity
™
Enteral Feeding Pump. Each alarm, message, or
indication is described in detail on the following pages:
Figure
Description
Page
Alarms
3-1
ERRA-ERRZ ...................................................................... 22
3-2
LOAD SET ........................................................................ 22
3-3
LOW BATT ........................................................................ 22
3-4
NO FLOW IN ..................................................................... 23
3-5
NO FLOW OUT ................................................................... 23
3-6
NO FOOD ......................................................................... 24
3-7
PUSH RUN TO FEED ............................................................ 25
Messages
3-8
DOSE DONE ...................................................................... 25
3-9
NEXT DOSE ....................................................................... 25
Indications
3-10
CHEK .............................................................................. 26
3-11
TO STOP LET GO ................................................................ 26
Battery Doesn’t Hold Its Charge ........................................... 27
Battery Doesn’t Charge ...................................................... 27
Charger Installed but No Plug Symbol Visible ......................... 27
Should any of these alarms or indications continue after troubleshooting, con-
tact your healthcare provider for pump service.
© Copyright 2004 ZEVEX Incorporated
3. Alarms, Messages, and Indications
21
3. ALARMS, MESSAGES, AND INDICATIONS
To adjust these settings:
Step 1:
With pump turned off, press and hold for 1.5 seconds the
+
and the
ON/OFF
keys at the same time. Pump will beep and then display the words BEEP
HIGH or BEEP LOW.
Press the
+
key to change to HIGH or the
-
key to change to LOW. Pump
will give a triple beep when the
+
or
-
keys are pressed and setting will
change. If setting is as desired, do not press either key.
Step 2:
Press the
PRIME
key. Display will read UNLK (unlocked) or LOCK (locked).
Press the
+
key to change to LOCK (locked) or the
-
key to change to UNLK
(unlocked). If setting is as desired, do not press either key.
Step 3:
Press the
PRIME
key again. Display will read MUTE WHEN DONE or BEEP
WHEN DONE.
Press the
+
key to change to BEEP WHEN DONE or the
-
key to change to
MUTE WHEN DONE. If setting is as desired, do not press either key.
Step 4:
Press the
PRIME
key again. Display will read LITE ON or LITE OFF.
Press the
+
key to change to LITE ON or the
-
key to change to LITE OFF.
If setting is as desired, do not press either key.
Step 5:
Press the
ON/OFF
to turn pump off. Setting changes will be saved auto-
matically.
Key Sequence for “Adjusting User Preference Settings”
HIGH
LOW
LOCK
UNLK
MUTE
BEEP
LITE ON
LITE OFF
25356_pages.qxd 2/2/2005 2:50 PM Page 26
To Purchase, Visit
or call