2. Directions for Use
12
EnteraLite
®
Infinity
™
Operator’s Manual
When you would like to restart pump, press the
ON/OFF
and/or the
RUN/PAUSE
key. Pump will save the memory of where it was in the feeding
before stopping. Press the
RUN/PAUSE
key and pump will restart at the
point where it was stopped.
NOTE: Any changes to pump settings during a feeding cycle will cause pump
to start a new feeding; it will not start where it left off.
Feeding Completion:
When the dose has been completely delivered, pump will stop running, and
display will read DOSE DONE. Pump will be in a pause mode until it is turned
off or feeding is restarted. To clear DOSE DONE, press and hold the
ON/OFF
key for 1.5 seconds to turn pump off, or press the
RUN/PAUSE
key to put
pump in pause mode. Adjustments to settings can be made at this time, or
press the
RUN/PAUSE
key a second time to start another feeding using
same rate, dose, and feed interval settings.
If you would prefer the alarm to sound when the dose is
complete, you can change the pump dose complete
alarm setting to BEEP WHEN DONE. See the User
Preference Settings section
(Page 19)
for instructions to
change this setting. When set to BEEP WHEN DONE, the
pump will stop at the end of the feeding, beep inter-
mittently, and display will read DOSE DONE (
Figure
2-16
). To silence alarm when the dose is complete,
press and hold the
ON/OFF
key for 1.5 seconds to turn
pump off, or press the
RUN/PAUSE
key to put pump in
pause mode.
NOTE: To program an infinite dose, press and hold the
+
key until display
reads INF. When an infinite dose has been programmed, pump will not alarm
DOSE DONE. Pump will run continuously as long as feeding solution remains
in the bag. When empty, pump will alarm NO FOOD or NO FLOW IN.
© Copyright 2004 ZEVEX Incorporated
29
Figure 2-16 DOSE DONE
5. ACCESSORIES
5. Accessories
Additional accessories are available for use with EnteraLite
®
Infinity
™
:
EnteraLite
®
Infinity
™
AC Adapter/Charger
Order Number INFCH01
Plug AC Adapter/Charger into a wall outlet and plug
connector into pump to operate on AC power
and/or recharge battery.
EnteraLite
®
Infinity
™
Pole Clamp
Order Number Z-11981
Thread bolt into receptacle on the back of pump
and tighten gray wheel to mount pump on clamp.
If gray wheel is slightly loosened, pump can be
rotated to snap into different positions. Retighten
gray wheel when pump is in the desired position.
Tighten black wheel to mount clamp on a pole.
Figure 5-1 AC Adapter/Charger
Figure 5-2 Pole Clamp
Figure 5-3 Pole Clamp with
Pump Attached to IV Pole