2. Directions for Use
7
2. DIRECTIONS FOR USE
Recommendation for First Use:
Since battery may not be fully charged when pump is first received, it is rec-
ommended that battery be charged for 6 hours prior to operating on battery
power
(see Page 18 for additional information).
Priming and Loading the
Disposable Set
Step 1:
If you are using the spike set with a vented
bottle or pre-filled bag:
Remove protective cover from spike, and insert
spike into spike adapter of formula container
(
Figure 2-1).
If you are using the 500ml or 1200ml dispos-
able set:
Hold bag upright and pour in feeding solution
(
Figure 2-2).
Close cap securely.
NOTE: Blenderized or aggressively mixed solu-
tions may have foam. If using this type of
solution, allow it to sit for 10 to 15 minutes
before pouring into bag. This will reduce the
chance of an alarm due to air in the tubing.
Step 2:
Remove protective cover from barbed adapter. If
using pole clamp, protective cover may be held
in the groove on back of clamp (
Figure 2-3).
Figure 2-1 Spike Container
Figure 2-2 Fill Bag
Figure 2-3 Remove Cover from
Barbed Adapter / Place Cover
on Pole Clamp
7. IMPORTANT PEDIATRIC CONSIDERATIONS
7. Important Pediatric Considerations
EnteraLite
®
Infinity
™
can be used on pediatric patients if the specifications of
pump meet delivery requirements of the patient. The specifications of primary
importance are:
The flow rate range of EnteraLite
®
Infinity
™
is 0.1 ml/hr to 10 ml/hr in 0.1
ml/hr increments and 10 ml/hr to 600 ml/hr in 1 ml/hr increments.
The EnteraLite
®
Infinity
™
pump delivers the dose at the specified rate with-
in +/- 5% volumetric flow rate accuracy under the following conditions:
NOTE: For test methodology refer to ANSI/AAMI ID26-1992 American National
Standard for Infusion Devices. Reported accuracy is based upon a generic vol-
ume accuracy calculation:
Accuracy % = (Volume Delivered - Volume Programmed) x 100
Volume Programmed
utilizing ZEVEX EnteraLite® Infinity™ disposable sets
fluid head height at +6.0 inches ± 0.3 inches (+15.24 cm ± 0.76 cm)
with respect to center of rotor.
over the range of environmental operating conditions (Specified per
IEC 60601-1)
Operating Temperature Range: 41 °F to 104 °F (5 °C to 40 °C)
Operating Humidity Range: 30 to 75% noncondensing