1. Pump Components
6
EnteraLite
®
Infinity
™
Operator’s Manual
Symbols
© Copyright 2004 ZEVEX Incorporated
35
Brief instructions for operation of pump as
well as a brief explanation of each alarm
message which pump may communicate are
printed on the label attached to pump.
These are not intended to be used in place
of the Operator's Manual. They are simply a
quick reference guide. Please read the
Operator's Manual before operating the
EnteraLite
®
Infinity
™
!
The bag symbol is printed on pump door
over port where upstream tubing enters
pump.
The patient symbol is printed on pump door
over port where downstream tubing exits
pump.
EN 60601-1 Type BF degree of protection
against electrical shock. No electrical con-
nection to patient. Drop from any angle
from height of 3 feet shall not damage
pump operation.
IEC 529 degree of protection against water
entering the enclosure. Water jets from any
direction shall have no effect.
This symbol is printed on pump delivery
sets. It indicates only one patient should
use each disposable set.
This symbol is printed on pump delivery
sets. It indicates pump delivery sets are
made with materials that do not contain the
plasticizer DEHP.
Figure 1-12
Feeding Bag Connection
Figure 1-13
Patient Connection
Figure 1-14
Type BF Shock Protection
Figure 1-15
IEC 529
IPX5
Figure 1-11
Consult Accompanying
Documents
Consult Accompanying
Documents!
Figure 1-16
Single Patient Use
Figure 1-17
DEHP-free
7. Important Pediatric Considerations
Occlusion alarm pressure of EnteraLite
®
Infinity
™
is 5 psi (34kPa) for
upstream occlusion and 12 psi (83 kPa) for a downstream occlusion.
If these specifications meet the required feed regimen, EnteraLite
®
Infinity
™
can be used to deliver enteral formulas to pediatric patients. ALWAYS VERIFY
RATE, DOSE AND FEED INTERVAL BEFORE PROCEEDING TO FEED.
25356_pages.qxd 2/2/2005 2:49 PM Page 12
To Purchase, Visit
or call