Operation
INTRABEAM Water Phantom
Version 6.2
Page 52
G-30-1767-en
The correlation between the two measurement methods is described by the
following formula:
Where f'(r) represents the conversion function. This formula has a standard
deviation of max. 5.1%. The values for the conversion function itself can be
obtained from the calibration documents of the XRS 4.
In summary, the conversion function comprises 3 different parts:
1
Use of two different ionization chamber models (PTW TN34013 vs.
TN23342) in the water phantom. The diameter of the chamber entrance
film of the active volume of the TN23342 is approx. 5.2 mm in contrast to
only 2.9 mm for the TN34013. This leads to different angular resolutions
and to different "true" distances r (typical geometric and absorption prob-
lem for point sources). Therefore, the smaller distance r is, the greater the
deviations between the two depth dose curves will be.
2
Use of two different designs for the ionization chamber holders. This leads
to different measured dose rates, which becomes especially noticeable for
small distances r.
3
Use of two different algorithms to calculate the water energy dose rate
(standard ion dose rate with a roentgen to gray conversion factor of 8.81
mGy/R in contrast to air kerma with the conversion factor
from
PTW), which leads to a constant offset in the entire measuring range.
Ḋ
w
(r){TARGIT} = f'(r) .
Ḋ
w
(r){Calibration v4.0}
K
Ka
D
W
Содержание INTRABEAM
Страница 1: ...ZEISS INTRABEAM Water Phantom Instructions for Use G 30 1767 en Version 6 2 2022 02 24...
Страница 4: ...INTRABEAM Water Phantom Version 6 2 Page 4 G 30 1767 en Blank page for your notes...
Страница 15: ...Version 6 2 G 30 1767 en Page 15 INTRABEAM Water Phantom Safety measures Blank page for your notes...
Страница 17: ...Version 6 2 G 30 1767 en Page 17 INTRABEAM Water Phantom Safety measures Fig 1 Symbols and labels 1 2 3...
Страница 28: ...Preparations for use INTRABEAM Water Phantom Version 6 2 Page 28 G 30 1767 en Blank page for your notes...
Страница 45: ...Version 6 2 G 30 1767 en Page 45 INTRABEAM Water Phantom Operation Blank page for your notes...
Страница 47: ...Version 6 2 G 30 1767 en Page 47 INTRABEAM Water Phantom Operation...
Страница 49: ...Version 6 2 G 30 1767 en Page 49 INTRABEAM Water Phantom Operation Blank page for your notes...
Страница 74: ...System data INTRABEAM Water Phantom Version 6 2 Page 74 G 30 1767 en Blank page for your notes...
Страница 80: ...Index INTRABEAM Water Phantom Version 6 2 Page 80 G 30 1767 en Blank page for your notes...
Страница 81: ...Version 6 2 G 30 1767 en Page 81 INTRABEAM Water Phantom Blank page for your notes...