Operation
INTRABEAM Water Phantom
Version 6.2
Page 30
G-30-1767-en
–
Make or disconnect the cable connections:
–
Operating the electrometer:
–
Remove any accumulated air bubbles from the X-ray probe.
–
Measure the water temperature and air pressure:
–
Measure the water temperature in the water tank.
–
Close the cover of the water tank properly.
–
Measure the air pressure.
–
Ensure that all covers are properly attached to the water tank.
–
Ensure that the XRS 4 is properly seated in the rotary feedthrough, i.e. that
there is no air gap between the XRS 4 and the XRS mount.
–
Ensure that the covers of the measuring chambers are properly attached.
–
Switch on the INTRABEAM and let it warm up.
–
Record the depth dose curve:
–
Place the probe tip in the optimum position (see page 35) in relation
to the IC ionization chamber in the measuring chamber for the depth
dose curve, check the positioning by measuring the current with the
electrometer.
–
Record the measured values. Depending on the number of measuring
points, this may take several hours.
–
Compare the measurement results with the stored calibration data of
the XRS 4.
–
Calculate the dose rate
–
Measure the isotropy:
INTRABEAM 600
•
Connect the IC ionization chamber to the “Ion Chamber” port.
•
Connect the XRS 4 to the XRS cable (green).
•
Open the folding shutter of the Workplace so that the electrometer is
accessible for operation.
INTRABEAM 600
(No action is required, the electrometer is automatically switched to
manual operation.)
Содержание INTRABEAM
Страница 1: ...ZEISS INTRABEAM Water Phantom Instructions for Use G 30 1767 en Version 6 2 2022 02 24...
Страница 4: ...INTRABEAM Water Phantom Version 6 2 Page 4 G 30 1767 en Blank page for your notes...
Страница 15: ...Version 6 2 G 30 1767 en Page 15 INTRABEAM Water Phantom Safety measures Blank page for your notes...
Страница 17: ...Version 6 2 G 30 1767 en Page 17 INTRABEAM Water Phantom Safety measures Fig 1 Symbols and labels 1 2 3...
Страница 28: ...Preparations for use INTRABEAM Water Phantom Version 6 2 Page 28 G 30 1767 en Blank page for your notes...
Страница 45: ...Version 6 2 G 30 1767 en Page 45 INTRABEAM Water Phantom Operation Blank page for your notes...
Страница 47: ...Version 6 2 G 30 1767 en Page 47 INTRABEAM Water Phantom Operation...
Страница 49: ...Version 6 2 G 30 1767 en Page 49 INTRABEAM Water Phantom Operation Blank page for your notes...
Страница 74: ...System data INTRABEAM Water Phantom Version 6 2 Page 74 G 30 1767 en Blank page for your notes...
Страница 80: ...Index INTRABEAM Water Phantom Version 6 2 Page 80 G 30 1767 en Blank page for your notes...
Страница 81: ...Version 6 2 G 30 1767 en Page 81 INTRABEAM Water Phantom Blank page for your notes...