Operation
INTRABEAM Water Phantom
Version 6.2
Page 34
G-30-1767-en
–
Tighten the clamping screw by hand to fix the rotary feedthrough in
one of its eight detent positions.
•
Establish all required cable connections:
•
Make the following settings to enable your INTRABEAM to emit X-rays:
•
Create a "patient", for example "XRS_test_1".
•
Make the settings for the XRS 4: 40
μ
A beam current, 50 kV acceleration
voltage, no applicator.
•
Select the following settings in order to obtain a treatment duration of
several hours: Prescribed dose 90 Gy and treatment depth 30 mm.
•
Start the "treatment" and wait until the internal X-ray monitor indicates
approx. 100%.
You will find detailed explanations on how to operate the measuring
functions in the Instructions for Use of your INTRABEAM (Instructions for Use
"INTRABEAM 600").
You can interrupt X-ray emission by pressing the <Pause> button.
You can continue X-ray emission by pressing the <Fortsetzen> button.
NOTE
If the probe tip is bent out of tolerance while being inserted in the water
phantom, the XRS 4 will no longer be able to emit radiation and the quality
assurance steps must be repeated.
•
Carefully insert the XRS 4 into the XRS mount in the rotary feedthrough.
•
Make sure that the probe tip does not touch the measuring chamber
while adjusting the positioning unit.
INTRABEAM 600
•
Connect the IC ionization chamber to the “Ion Chamber” port.
•
Connect the XRS 4 to the XRS cable (green).
INTRABEAM 600
•
Call up the “Treatment prescription” window
Содержание INTRABEAM
Страница 1: ...ZEISS INTRABEAM Water Phantom Instructions for Use G 30 1767 en Version 6 2 2022 02 24...
Страница 4: ...INTRABEAM Water Phantom Version 6 2 Page 4 G 30 1767 en Blank page for your notes...
Страница 15: ...Version 6 2 G 30 1767 en Page 15 INTRABEAM Water Phantom Safety measures Blank page for your notes...
Страница 17: ...Version 6 2 G 30 1767 en Page 17 INTRABEAM Water Phantom Safety measures Fig 1 Symbols and labels 1 2 3...
Страница 28: ...Preparations for use INTRABEAM Water Phantom Version 6 2 Page 28 G 30 1767 en Blank page for your notes...
Страница 45: ...Version 6 2 G 30 1767 en Page 45 INTRABEAM Water Phantom Operation Blank page for your notes...
Страница 47: ...Version 6 2 G 30 1767 en Page 47 INTRABEAM Water Phantom Operation...
Страница 49: ...Version 6 2 G 30 1767 en Page 49 INTRABEAM Water Phantom Operation Blank page for your notes...
Страница 74: ...System data INTRABEAM Water Phantom Version 6 2 Page 74 G 30 1767 en Blank page for your notes...
Страница 80: ...Index INTRABEAM Water Phantom Version 6 2 Page 80 G 30 1767 en Blank page for your notes...
Страница 81: ...Version 6 2 G 30 1767 en Page 81 INTRABEAM Water Phantom Blank page for your notes...