Operation
INTRABEAM Water Phantom
Version 6.2
Page 60
G-30-1767-en
•
Observe the current value reading on the electrometer display and move
the spherical applicator further upwards. You can see that the current
value increases accordingly.
•
If you move the spherical applicator together with the X-ray probe further
upwards and the current value drops, you have passed the exact target
point.
•
Move the spherical applicator together with the X-ray probe back down
slightly and vary your setting in the Z axis until you have set the maximum
current value.
•
In the position with the maximum current value, you have reached the ex-
act Z position of the spherical applicator (and of the probe tip).
•
Secure this Z positioning using the locking lever on the adjusting knob for
the Z axis.
The isotropy measurement is performed by rotating the XRS 4 through one
full revolution while measuring the emitted X-ray radiation.
•
Position the rotary feedthrough in the 8 detent positions.
The procedure for positioning the rotary feedthrough is explained on
page 57.
•
Determine the isotropy and calculate the deviations.
The procedure for determining the isotropy and calculating the deviations
is explained on page 58.
Recording the depth dose curve with a flat applicator or
surface applicator
Exact positioning of the applicator (with the probe tip) is a precondition for
accurate and meaningful measuring results.
•
Do not adjust the positioning unit in the X and Y axes.
•
Unless this has already been done:Note down the previously determined
value of Z
1
•
Release the locking lever on the adjusting knob for the Z axis.
•
Place the positioning unit all the way up in the Z axis.
Содержание INTRABEAM
Страница 1: ...ZEISS INTRABEAM Water Phantom Instructions for Use G 30 1767 en Version 6 2 2022 02 24...
Страница 4: ...INTRABEAM Water Phantom Version 6 2 Page 4 G 30 1767 en Blank page for your notes...
Страница 15: ...Version 6 2 G 30 1767 en Page 15 INTRABEAM Water Phantom Safety measures Blank page for your notes...
Страница 17: ...Version 6 2 G 30 1767 en Page 17 INTRABEAM Water Phantom Safety measures Fig 1 Symbols and labels 1 2 3...
Страница 28: ...Preparations for use INTRABEAM Water Phantom Version 6 2 Page 28 G 30 1767 en Blank page for your notes...
Страница 45: ...Version 6 2 G 30 1767 en Page 45 INTRABEAM Water Phantom Operation Blank page for your notes...
Страница 47: ...Version 6 2 G 30 1767 en Page 47 INTRABEAM Water Phantom Operation...
Страница 49: ...Version 6 2 G 30 1767 en Page 49 INTRABEAM Water Phantom Operation Blank page for your notes...
Страница 74: ...System data INTRABEAM Water Phantom Version 6 2 Page 74 G 30 1767 en Blank page for your notes...
Страница 80: ...Index INTRABEAM Water Phantom Version 6 2 Page 80 G 30 1767 en Blank page for your notes...
Страница 81: ...Version 6 2 G 30 1767 en Page 81 INTRABEAM Water Phantom Blank page for your notes...