Operation
INTRABEAM Water Phantom
Version 6.2
Page 62
G-30-1767-en
Water phantom procedure for depth dose curve measurements using a
flat or surface applicator
The applicators are designed for radiation treatment on the surface (surface
applicator) and at a depth of 5 mm (flat applicator). In the first case, the dose
rate must be measured on the surface of the applicator. The ionization
chamber itself is not watertight, so it has to be maintained watertight in a
measuring chamber. Because of the wall of the measuring chamber, the ap-
plicators cannot be placed flat directly in the ionization chamber.
This is a problem similar to that of spherical applicators when the dose rate
has to be measured there on the applicator surface. For this reason, an adjust-
ment function must be applied to the depth dose curve or the transfer func-
tion to extrapolate the data to the surface of the spherical applicators. The
same applies to the flat or surface applicators. Therefore, in order to simulate
the clinical conditions, the measuring chamber should touch the surface of
the applicator placed in a water/air interface completely as shown in the illus-
tration below.
Chamber holder making contact with the surface of the
applicator placed flush in water (air-water interface)
Содержание INTRABEAM
Страница 1: ...ZEISS INTRABEAM Water Phantom Instructions for Use G 30 1767 en Version 6 2 2022 02 24...
Страница 4: ...INTRABEAM Water Phantom Version 6 2 Page 4 G 30 1767 en Blank page for your notes...
Страница 15: ...Version 6 2 G 30 1767 en Page 15 INTRABEAM Water Phantom Safety measures Blank page for your notes...
Страница 17: ...Version 6 2 G 30 1767 en Page 17 INTRABEAM Water Phantom Safety measures Fig 1 Symbols and labels 1 2 3...
Страница 28: ...Preparations for use INTRABEAM Water Phantom Version 6 2 Page 28 G 30 1767 en Blank page for your notes...
Страница 45: ...Version 6 2 G 30 1767 en Page 45 INTRABEAM Water Phantom Operation Blank page for your notes...
Страница 47: ...Version 6 2 G 30 1767 en Page 47 INTRABEAM Water Phantom Operation...
Страница 49: ...Version 6 2 G 30 1767 en Page 49 INTRABEAM Water Phantom Operation Blank page for your notes...
Страница 74: ...System data INTRABEAM Water Phantom Version 6 2 Page 74 G 30 1767 en Blank page for your notes...
Страница 80: ...Index INTRABEAM Water Phantom Version 6 2 Page 80 G 30 1767 en Blank page for your notes...
Страница 81: ...Version 6 2 G 30 1767 en Page 81 INTRABEAM Water Phantom Blank page for your notes...