Operation
INTRABEAM Water Phantom
Version 6.2
Page 38
G-30-1767-en
•
Fix the setting of the Z position by actuating the locking lever on the ad-
justing knob for the Z axis.
•
Search for the exact center position for the probe tip via the measuring
window in the X and Y coordinate axes in succession:
1
2
NOTE
If the probe tip is bent out of tolerance while being inserted in the water
phantom, the XRS 4 will no longer be able to emit radiation and the quality
assurance steps must be repeated.
•
Carefully insert the XRS 4 into the XRS mount in the rotary feedthrough.
•
Make sure that the probe tip does not touch the measuring chamber
while adjusting the positioning unit.
Содержание INTRABEAM
Страница 1: ...ZEISS INTRABEAM Water Phantom Instructions for Use G 30 1767 en Version 6 2 2022 02 24...
Страница 4: ...INTRABEAM Water Phantom Version 6 2 Page 4 G 30 1767 en Blank page for your notes...
Страница 15: ...Version 6 2 G 30 1767 en Page 15 INTRABEAM Water Phantom Safety measures Blank page for your notes...
Страница 17: ...Version 6 2 G 30 1767 en Page 17 INTRABEAM Water Phantom Safety measures Fig 1 Symbols and labels 1 2 3...
Страница 28: ...Preparations for use INTRABEAM Water Phantom Version 6 2 Page 28 G 30 1767 en Blank page for your notes...
Страница 45: ...Version 6 2 G 30 1767 en Page 45 INTRABEAM Water Phantom Operation Blank page for your notes...
Страница 47: ...Version 6 2 G 30 1767 en Page 47 INTRABEAM Water Phantom Operation...
Страница 49: ...Version 6 2 G 30 1767 en Page 49 INTRABEAM Water Phantom Operation Blank page for your notes...
Страница 74: ...System data INTRABEAM Water Phantom Version 6 2 Page 74 G 30 1767 en Blank page for your notes...
Страница 80: ...Index INTRABEAM Water Phantom Version 6 2 Page 80 G 30 1767 en Blank page for your notes...
Страница 81: ...Version 6 2 G 30 1767 en Page 81 INTRABEAM Water Phantom Blank page for your notes...