Version 6.2
G-30-1767-en
Page 31
INTRABEAM Water Phantom
Operation
–
Place the probe tip in the optimum position in relation to the IC
ionization chamber in the measuring chamber for isotropy, i.e. only
raise it along the Z axis; check the positioning by measuring the
current with the electrometer.
–
Record the measured values. To do this, use the rotary feedthrough to
rotate the XRS 4 to the 8 detent positions, fixing it in the detent posi-
tion with the clamping screw in each case.
–
Calculate the deviation from the ideal isotropy.
–
Drain the water from the water tank.
The protective CrN coating of the X-ray probe can be damaged.
•
Do not leave the X-ray probe of the XRS 4 in liquid (e.g. water, cleaning
agents or disinfectants) for more than 2 hours. Failure to comply with this
instruction may result in damage to the protective CrN coating of the X-
ray probe.
Содержание INTRABEAM
Страница 1: ...ZEISS INTRABEAM Water Phantom Instructions for Use G 30 1767 en Version 6 2 2022 02 24...
Страница 4: ...INTRABEAM Water Phantom Version 6 2 Page 4 G 30 1767 en Blank page for your notes...
Страница 15: ...Version 6 2 G 30 1767 en Page 15 INTRABEAM Water Phantom Safety measures Blank page for your notes...
Страница 17: ...Version 6 2 G 30 1767 en Page 17 INTRABEAM Water Phantom Safety measures Fig 1 Symbols and labels 1 2 3...
Страница 28: ...Preparations for use INTRABEAM Water Phantom Version 6 2 Page 28 G 30 1767 en Blank page for your notes...
Страница 45: ...Version 6 2 G 30 1767 en Page 45 INTRABEAM Water Phantom Operation Blank page for your notes...
Страница 47: ...Version 6 2 G 30 1767 en Page 47 INTRABEAM Water Phantom Operation...
Страница 49: ...Version 6 2 G 30 1767 en Page 49 INTRABEAM Water Phantom Operation Blank page for your notes...
Страница 74: ...System data INTRABEAM Water Phantom Version 6 2 Page 74 G 30 1767 en Blank page for your notes...
Страница 80: ...Index INTRABEAM Water Phantom Version 6 2 Page 80 G 30 1767 en Blank page for your notes...
Страница 81: ...Version 6 2 G 30 1767 en Page 81 INTRABEAM Water Phantom Blank page for your notes...