Conclusions
The grayish white color first occurring in the bright position in the above example corresponds to a path
difference of 150 nm, according to the Michel Lévy color chart (3-23).
The "surrounding" of the artificial fiber, which is not birefringent, displays a dark red when the
λ
-
compensator is inserted, which corresponds to the path difference of the compensator of 550 nm
(1
st
order interference color for the path difference 550 nm, corresponds to 1
λ
).
If the vibration direction of the birefringent specimen to be examined is parallel to the largest vibration
direction of the
λ
-compensator, i.e. in NE-SW direction, the path difference of the specimen (e.g. grayish
white: 150 nm) and the path difference of the
λ
-compensator (red: 550 nm) add up. This results in a
color change of the specimen from grayish white to greenish blue (resulting path difference = 700 nm).
If the vibration direction of the specimen to be examined is perpendicular to the largest vibration
direction of the
λ
-compensator, i.e. in NW-SE direction, the path difference of the specimen (e.g. grayish
white: 150 nm) is subtracted from the path difference of the
λ
-compensator (red: 550 nm). A change in
the interference color of the specimen from grayish white to orange (resulting path difference = 400 nm)
then becomes visible.
☞
Michel Lévy color charts are available under order No. 42-312.
Содержание Axioskop 2
Страница 1: ...Axioskop 2 Routine Microscope Operating Manual...
Страница 14: ...INTRODUCTION Axioskop 2 Axioskop 2 MOT 0 14 B 40 075 e 06 98...
Страница 120: ...OPERATION Axioskop 2 Axioskop 2 MOT 3 54 B 40 075 e 06 98...
Страница 122: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 2 Axioskop 2 MOT 4 2 B 40 075 e 06 98...
Страница 132: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 2 Axioskop 2 MOT 4 12 B 40 075 e 06 98...
Страница 134: ...ANNEX Axioskop 2 Axioskop 2 MOT A 2 B 40 075 e 06 98...