Attachment of universal condenser
•
Move stage carrier to the upper stop position
via the focusing drive.
•
Use lever (2-7/
8
) to fold out the front lens on
the condenser or, in the case of the motorized
version, swing it out by pressing the respective
key. Unscrew both centering screws (2-7/
3
)
until their ends are no longer visible.
•
Move condenser carrier (2-7/
2
) to the lowest
position via drive for height adjustment (2-7/
5
).
•
Insert condenser (2-7/
1
) between condenser
carrier (2-7/
2
) and stage carrier (2-7/
4
). Orient
the locking screw on the underside of the
condenser in the direction of the groove
(2-7/
7
).
•
Press dovetail of condenser against the spring
mount (2-7/
6
) of the condenser carrier until the
condenser can be placed on the condenser
carrier horizontally.
•
Let condenser go smoothly; the screw will lock
in position in the groove (2-7/
7
) at the front.
•
Screw in centering screws until they engage in
the dovetail of the condenser.
•
In the case of the motorized condenser
(Axioskop 2
MOT
), plug connecting cable in the
socket on the left front of the stand (2-8/
1
) and
tighten screwing ring.
☞
The achromatic-aplanatic condenser
1.4 H DIC and other condensers
from the Axioplan 2 line are also
suitable for attachment.
Fig. 2-7
Attachment of universal condenser
Fig. 2-8
Attachment of universal condenser
MOT
Содержание Axioskop 2
Страница 1: ...Axioskop 2 Routine Microscope Operating Manual...
Страница 14: ...INTRODUCTION Axioskop 2 Axioskop 2 MOT 0 14 B 40 075 e 06 98...
Страница 120: ...OPERATION Axioskop 2 Axioskop 2 MOT 3 54 B 40 075 e 06 98...
Страница 122: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 2 Axioskop 2 MOT 4 2 B 40 075 e 06 98...
Страница 132: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 2 Axioskop 2 MOT 4 12 B 40 075 e 06 98...
Страница 134: ...ANNEX Axioskop 2 Axioskop 2 MOT A 2 B 40 075 e 06 98...