![Zeiss Axioskop 2 Скачать руководство пользователя страница 52](http://html1.mh-extra.com/html/zeiss/axioskop-2/axioskop-2_manual_3373791052.webp)
START-UP
Attachments and conversions
Axioskop 2 / Axioskop 2
MOT
2-20
B 40-075 e 06/98
After expiry of the average operation time of
300 h, the HBO 103 W/2 mercury vapor short-arc
lamp must be exchanged.
The operating time of the lamp can be checked on
the counter (2-30/
1
) of the HBO
100
W
transformer.
Follow the following steps when changing the
HBO 103 W/2 lamp:
•
Remove or unscrew the line plug and the HBO
103 plug from the HBO 100 W transformer.
•
Use SW 3 ball-headed screwdriver to loosen
clamping screw (2-31/
2
) and remove HBO 103
illuminator (2-31/
3
) from the incident-light tube
(2-31/
1
) of the microscope stand.
•
Wear protective mask and safety gloves.
•
Use focusing knob (2-32/
1
) to bring collector to
the position at the very front (in the direction of
the light).
•
Use ball-headed screwdriver to loosen clamping
screw for lamp housing (2-32/
11
) and carefully
remove lamp housing (2-32/
2
) from the lamp
mount (2-32/
9
) in an upward direction.
•
Use knurled screw (2-32/
12
) to loosen the
connection between contact wire and ceramic
body.
•
Use 17 mm fork wrench (2-32/
6
) to carefully
loosen hexagonal nut (2-32/
7
). Take out
HBO
103 W/2 lamp (2-32/
5
) including the
cooling body (2-32/
3
) and carefully place the
unit on the table with the cooling body
pointing downwards.
•
Loosen the clamping screw (2-32/
4
) on the
cooling body, carefully remove the HBO 103
W/2 lamp (2-32/
5
) from the cooling body and
dispose of it in the original packaging in
accordance with the legal regulations.
Fig. 2-30
HBO 100 W transformer
Fig. 2-31
Removing the HBO 103 illuminator
Содержание Axioskop 2
Страница 1: ...Axioskop 2 Routine Microscope Operating Manual...
Страница 14: ...INTRODUCTION Axioskop 2 Axioskop 2 MOT 0 14 B 40 075 e 06 98...
Страница 120: ...OPERATION Axioskop 2 Axioskop 2 MOT 3 54 B 40 075 e 06 98...
Страница 122: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 2 Axioskop 2 MOT 4 2 B 40 075 e 06 98...
Страница 132: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 2 Axioskop 2 MOT 4 12 B 40 075 e 06 98...
Страница 134: ...ANNEX Axioskop 2 Axioskop 2 MOT A 2 B 40 075 e 06 98...