START-UP
Attachments and conversions
Axioskop 2 / Axioskop 2
MOT
2-24
B 40-075 e 06/98
(3)
Setting the smoothness (torque) of the
drive for stage adjustment
The torque for x and y movement can be
individually matched for both mechanical stages.
They are factory-aligned to a medium value.
a)
X-drive
•
Insert SW 2 mm screwdriver in the drilled hole
of nut (2-37/
3
).
•
Hold drive for x-movement (2-37/
4
) and use
screwdriver to set the nut clockwise (greater
torque) or anti-clockwise (less torque) until the
required smoothness is obtained.
b)
Y-drive
•
Insert supplied SW 2 mm screwdriver in the
drilled hole of drive (2-37/
2
).
•
Turn drive (2-37/
1
) to the right or left until the
screwdriver clickstops in one of the holes of the
inside nut.
•
Hold drive for y-movement (2-37/
1
) and use inserted screwdriver to set the nut clockwise (less torque)
or anti-clockwise (greater torque) until the required smoothness is obtained.
☞
The setting range of the torque for the x and y drives amounts to approx. ¾ rotations each.
Further movement in the direction of less torque will not result in a reduction of torque and
should be avoided.
To safeguard a long life of the stage, the scuff from the specimen holders should be removed at
regular intervals. Make sure that the scuff does not enter the guiding components for the x-
movement.
X
Y
Fig. 2-37
Setting the torque of the coaxial
stage drive
Содержание Axioskop 2
Страница 1: ...Axioskop 2 Routine Microscope Operating Manual...
Страница 14: ...INTRODUCTION Axioskop 2 Axioskop 2 MOT 0 14 B 40 075 e 06 98...
Страница 120: ...OPERATION Axioskop 2 Axioskop 2 MOT 3 54 B 40 075 e 06 98...
Страница 122: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 2 Axioskop 2 MOT 4 2 B 40 075 e 06 98...
Страница 132: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 2 Axioskop 2 MOT 4 12 B 40 075 e 06 98...
Страница 134: ...ANNEX Axioskop 2 Axioskop 2 MOT A 2 B 40 075 e 06 98...