START-UP
Axioskop 2 / Axioskop 2
MOT
Attachments and conversions
B 40-075 e 06/98
2-25
(4)
Changing the specimen holder
•
Loosen two countersunk screws (2-38/
1
) using
SW 2 Allen key (2-38/
3
).
•
Remove specimen holder (2-38/
2
) from the
mechanical stage (2-38/
4
) in an upward
direction.
•
Attach required specimen holder (2-38/
5
) to the
mechanical stage and tighten two countersunk
screws.
2.2.9
Changing the binocular tube
•
Use SW 3 ball-headed screwdriver to loosen
clamping screw (2-39/
3
) and remove available
tube in an upward direction.
•
Place dust cap (2-39/
2
) for tube lens protection
above the dovetail of the binocular tube.
•
Remove dust cap from the required tube.
•
Insert dovetail of tube (2-39/
1
) in the stand
opening (2-39/
4
) and align the tube.
•
Tighten clamping screw (2-39/
3
).
2
1
3
5
4
Fig. 2-38
Changing the specimen holder
4
3
2
1
2
1
4
3
Fig. 2-39
Changing the binocular tube
Содержание Axioskop 2
Страница 1: ...Axioskop 2 Routine Microscope Operating Manual...
Страница 14: ...INTRODUCTION Axioskop 2 Axioskop 2 MOT 0 14 B 40 075 e 06 98...
Страница 120: ...OPERATION Axioskop 2 Axioskop 2 MOT 3 54 B 40 075 e 06 98...
Страница 122: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 2 Axioskop 2 MOT 4 2 B 40 075 e 06 98...
Страница 132: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 2 Axioskop 2 MOT 4 12 B 40 075 e 06 98...
Страница 134: ...ANNEX Axioskop 2 Axioskop 2 MOT A 2 B 40 075 e 06 98...