START-UP
Axioskop 2 / Axioskop 2
MOT
Attachments and conversions
B 40-075 e 06/98
2-17
2.2.3
Changing the DIC prism in the
universal condenser
•
To loosen the cover on the condenser underside
(2-26/
1
), loosen both grub screws using the
SW
2
screwdriver and remove the cover.
Position the turret disk containing the DIC prism
to be exchanged in the exchange opening and
hold it on the knurled ring.
•
Unscrew retainer ring using the so-called
double-function tool (2-26/
4
).
•
Now screw the tool in the prism mount (2-26/
3
)
with the threaded side (2-26/
4
) and pull out
DIC prism (2-26/
2
).
•
Remove DIC prism (2-26/
2
) from the tool and
screw on the new, required DIC prism.
•
Installation of the DIC prism is made in reverse
order. Make sure that the DIC prism is oriented
correctly (groove 2-26/
5
must engage in the
pinion of the mount). Make sure that the
knurled ring of the turret disk is labelled
correctly.