CARE, MAINTENANCE, TROUBLESHOOTING AND SERVICE
Axioskop 2 / Axioskop 2
MOT
Troubleshooting
B 40-075 e 06/98
4-7
Problem
Cause
Remedy
Low resolving power and poor
image contrast.
Aperture diaphragm opening not
set correctly.
Set aperture diaphragm opening
in accordance with the 2/3 rule or
depending on the specimen
features, see p. 3-24 ff.
Condenser not focused correctly
and condenser lens 0.9 not
switched correctly.
Focus condenser and switch
condenser lens 0.9 on or off
correctly, see p. 3-24 ff.
Wrong cover slip thickness for
transmitted-light objectives with
cover slip thickness 0.17 mm.
Use standardized cover slips with
thickness 0.17 mm.
Use of no or unspecified
immersion oil with immersion
objectives.
Use immersion oil 518 N from Carl
Zeiss, see p. 4-10.
Air bubbles in the immersion oil.
Apply new oil to remove the
bubbles.
Immersion oil at the front lens of a
dry objective.
Clean the front lens of the dry
objective, see p. 4-3.
Corr. ring is not set to the correct
cover slip thickness.
Set the corr. ring to the correct
thickness, see p. 1-12.
Dirt or dust on the optical surfaces
of objectives, eyepieces,
condensers or filters.
Clean the appropriate optical
components, see p. 4-3.
Asymmetrically blurred images,
e.g. one side in focus, one side
Condenser not set correctly.
Set condenser correctly,
see p. 3-24 ff.
not in focus.
Nosepiece not in click-stop
position.
Correctly click-stop nosepiece.
Specimen is not clamped in
position on the mechanical stage.
Correctly insert specimen in
specimen holder and clamp it.
Содержание Axioskop 2
Страница 1: ...Axioskop 2 Routine Microscope Operating Manual...
Страница 14: ...INTRODUCTION Axioskop 2 Axioskop 2 MOT 0 14 B 40 075 e 06 98...
Страница 120: ...OPERATION Axioskop 2 Axioskop 2 MOT 3 54 B 40 075 e 06 98...
Страница 122: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 2 Axioskop 2 MOT 4 2 B 40 075 e 06 98...
Страница 132: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 2 Axioskop 2 MOT 4 12 B 40 075 e 06 98...
Страница 134: ...ANNEX Axioskop 2 Axioskop 2 MOT A 2 B 40 075 e 06 98...