START-UP
Axioskop 2 FS
MOT
Initial start-up
Carl Zeiss
B 40-076 e 02/01
2-15
Connect HBO 103 illuminator and switch it
on/off
The HBO 103 illuminator (for epi-fluorescence) is
supplied via a separate transformer and switched
on/off via the power unit of this transformer.
•
Insert multi-pin plug of the HBO 103 illuminator
into the instrument socket (2-16/
2
) of the
HBO 100 W transformer and secure it with
captive coupling ring (also see 2.1.14).
•
The HBO
100
W transformer must be
connected to the line via its line socket
(2-16/
3
).
•
Switch the HBO 100 W transformer on/off
using line switch (2-16/
1)
.
Fig. 2-16
HBO 100 W transformer
(front and rear)