START-UP
Axioskop 2 FS
plus
Carl Zeiss
Initial start-up
Axioskop 2 FS
MOT
2-8
B 40-076 e 02/01
2.1.5
Screw in objectives
•
Remove the stage carrier (see 2.1.4) and place
it on the table in such a way that toppling is
prevented.
•
Move front objective mount (2-5/
2
) into the
work position via the pushrod (2-5/
1
) (also see
page 3-8).
•
Remove objective (2-5/
3
) with the required
magnification factor from the case and screw it
in the front mount of the objective slider until
stop.
•
Move rear objective mount (2-5/
4
) into the
work position via pushrod (2-5/
1
). Remove the
second objective (2-5/
5
) from the case and
screw it into the rear mount until stop.
☞
Attach stage carrier again only after inserting the DIC slider (if required) and the condenser.
Fig. 2-5
Screwing in of objectives