OPERATION
Axioskop 2 FS
plus
Carl Zeiss
Illumination and contrasting techniques
Axioskop 2 FS
MOT
3-20
B 40-076 e 02/01
3.3
Illumination and contrasting techniques
3.3.1
Setting of transmitted-light brightfield according to KÖHLER
(1)
General principle
Transmitted-light brightfield microscopy is the most usual of all the optical techniques, since it allows the
easy and fast viewing of high-contrast or stained specimens (e.g. blood smears).
For as true-to-object imaging as possible, indirect ray bundles, i.e. ray bundles diffracted and scattered
on the specimen details, are of major importance in addition to the so-called direct ray bundles. The
higher the portion of indirect rays (aperture), the more realistic the microscope image will be, according
to ABBE's rule.
To utilize the entire optical performance of the microscope, especially of the objective, the condenser,
the luminous-field diaphragm and the aperture diaphragm should be set in accordance with the rules for
KÖHLER illumination. These basic rules of microscope setting are described in detail in section 3.3.1 (3)
"Transmitted-light brightfield settings according to KÖHLER on the manual Axioskop 2 FS
plus
“ .
(2)
Transmitted-light brightfield equipment of the manual Axioskop 2 FS plus
Each manual Axioskop 2 FS
plus
configuration permits performance of the transmitted-light brightfield
technique.
(3)
Transmitted-light brightfield settings according to KÖHLER on the manual
Axioskop 2 FS plus
−
The manual Axioskop 2 FS
plus
has been set up as described in chapter 2.
−
Switch on the transmitted-light illumination via the separate power unit.
•
Set the image brightness via the light intensity control (3-1/
24
) of the separate power unit.
•
Place a high-contrast specimen in the universal mounting frame of the mechanical stage.
•
Swing in front lens of the universal condenser (3-16/
2
) (for objectives
≥
10x) and use drive for height
adjustment (3-17/
2
) to move the condenser to the upper stop. The stop must be set in such a way
that the specimen is not touched by the condenser (for setting of the condenser stop see section
3.3.1
(4)).
•
Switch turret disk of condenser (if available) to position H for brightfield via the knurled ring. If no
separate brightfield position is available, the turret position (I, II, or III) can be chosen which does not
contain a DIC prism. If all three positions are equipped with DIC prisms, switch off the analyzer in the
reflector turret.
☞
One turret position of the universal condenser should remain empty for highest demands in
transmitted-light brightfield.
Содержание Axioskop 2 FS plus
Страница 1: ...Operating Manual Axioskop 2 FS plus FS MOT Research Microscope...
Страница 14: ...INTRODUCTION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 0 14 B 40 076 e 02 01...
Страница 36: ...START UP Axioskop 2 FS plus Carl Zeiss List of illustrations Axioskop 2 FS MOT 2 4 B 40 076 e 02 01...
Страница 110: ...OPERATION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 3 40 B 40 076 e 02 01...
Страница 112: ...CARE MAINTENANCE TROUBLESHOOTING Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 4 2 B 40 076 e 02 01...
Страница 122: ...ANNEX Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT A 2 B 40 076 e 02 01...