OPERATION
Axioskop 2 FS
plus
Carl Zeiss
Operation and function controls
Axioskop 2 FS
MOT
3-18
B 40-076 e 02/01
b) Perform
steps
•
Each press of the
Step
key moves the objective slider one step further in the set direction.
•
Press the
Step L/H
key to set the speed in the Step mode to slow (L = low) or to maximum (H =
high). When maximum speed has been selected, the LED of the
Step L/H
key
lights up.
End switches of the work range
The software integrated in the control unit permits a work range for objective focusing to be defined via
electronic end switches. When these end switches are reached, the focusing drive is switched off
automatically.
The work range can be defined as required in the entire travel range of the objective slider, and activated
or deactivated any time.
☞
If the end switches for the work range are activated, the Home function is not available.
a)
Set end switches
•
Use the
direction key (or the handwheel) to approach the lowest position of the work range to be
defined.
•
Press the
Set End Switch
and the
direction key simultaneously to set the lower end switch.
•
Use the
direction key (or the handwheel) to approach the uppermost position of the work range to
be defined.
•
Press the
Set End Switch
and the
direction key simultaneously to set the upper end switch.
b)
Activate / deactivate end switches
•
Press the
End Switch on/off
key and the
direction key simultaneously to activate the end
switches. The LED of the
End Switch on/off
key lights up.
•
Press the
End Switch on/off
key and the
direction key simultaneously to deactivate the end
switches. You can now go beyond the limits of the defined work range within the mechanical travel
range.
Содержание Axioskop 2 FS plus
Страница 1: ...Operating Manual Axioskop 2 FS plus FS MOT Research Microscope...
Страница 14: ...INTRODUCTION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 0 14 B 40 076 e 02 01...
Страница 36: ...START UP Axioskop 2 FS plus Carl Zeiss List of illustrations Axioskop 2 FS MOT 2 4 B 40 076 e 02 01...
Страница 110: ...OPERATION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 3 40 B 40 076 e 02 01...
Страница 112: ...CARE MAINTENANCE TROUBLESHOOTING Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 4 2 B 40 076 e 02 01...
Страница 122: ...ANNEX Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT A 2 B 40 076 e 02 01...