OPERATION
Axioscope 5/7/Vario
Illumination and contrast methods in transmitted light …
ZEISS
03/2018
430035-7344-001
89
(4) Adjusting the height stop on the
condenser carrier
•
Loosen the setscrew (
1
) on the height
stop with the Allen wrench (AF 3).
•
Focus the sample with the focusing drive.
•
Close the field diaphragm and adjust the
condenser vertically (
2
) until you get a
sharp image.
•
Perform the vertical adjustment on the
condenser by small increments only to avoid
lifting out the sample.
•
Tighten the setscrew (
1
) on the height
stop.
(5) Adjusting the height stop on the focusing
drive
•
Turn the clamping lever (
1
) of the stop
towards the pin stop.
•
Move the stage with the aid of the focusing
drive (
2
) to the highest position
allowable without risking colliding with
specimen holder or objectives.
•
Now press the clamping lever down until the
stop engages again.
Fig. 4-4
Adjusting the height stop on the
condenser carrier
Fig. 4-5
Adjusting the height stop on the
focusing drive