CARE, FUSE REPLACEMENT AND SERVICE
Axioscope 5/7/Vario
Instrument maintenance
ZEISS
03/2018
430035-7344-001
127
5.2.3
Replacing the fuses in the
12 VDC 100 W power supply unit
CAUTION
Disconnect the instrument from the
power supply before handling the
instrument!
If fuses blow, the cause must be identified and any
technical defect must be properly repaired.
The fuse compartment is on the back of the power
supply unit and contains two fuses of the type
T 5 A/H 250 V, 5x20 mm
.
•
Disconnect the power plug.
•
Pull the fuse retainer (Fig. 5-2/
2
) out to the
front with the aid of a small screwdriver, if
necessary.
•
Remove the fuses from their retainer and
replace them with new fuses.
•
Push the fuse retainer back into the
compartment (Fig. 5-2/
1
) to the stop.
•
Insert the power plug again.
Fig. 5-2
Replacing fuses in the power supply
unit