START-UP
ZEISS
Mounting standard components
Axioscope 5/7/Vario
68
430035-7344-001
03/2018
3.1.21.3
Adjusting the HBO 100
illuminator
The HBO 100 illuminator is available in two
versions (manual and automatic adjustment).
The self-adjusting HBO 100 (423011-9901-000)
adjusts automatically when the power supply unit
is turned on.
Follow the instructions below to adjust the
HBO 100 illuminator manually (423010-0000-000).
NOTE
If the FL attenuator is in the reflected
light beam path, make sure that it is set
to 100% transmission.
•
Turn on the HBO 100 illuminator (
1
)
on the power supply unit for the HBO 100
(
1
) and wait until it reaches its
operational temperature.
•
Pull the adjustment tool (
1
) out of the
microscope stand. The lighter colored focal
point of the lamp HBO 103 W/2 and its slightly
darker reflection become visible in the black
glass window of the adjustment tool.
•
Use the knurled button (
4
) for
collector adjustment to focus the brighter focal
point.
•
Use the adjustment screws (
2
and
3
) to focus the darker focal point (reflection) as you would
for the focal point image (
2
) in the inscribed adjustment circle.
NOTE
The two focal points of the HBO 103 W/2 lamp should be close to each other in the
adjustment circle of the adjustment tool!
•
Replace the adjustment tool in its original position.
Fig. 3-35
Adjustment tool
Fig. 3-36
Adjusting the HBO 100