START-UP
Axioscope 5/7/Vario
Mounting standard components
ZEISS
03/2018
430035-7344-001
49
b) How to mount the stage
•
Fit the stage (
1
) onto the stage carrier (
6
) so that the threaded holes in the bottom
of the stage (
2
) are on top of the corresponding holes in the stage carrier (
3
).
•
Insert the four mounting screws (
5
) from below through the holes in the stage carrier and
screw them into the bottom of the stage.
•
Tighten the screws.
NOTE
After long hours of use, the stage guide may be loose and the stage travel range may be
decreased. However, this is not a malfunction and can be corrected easily as described below.
c) How to adjust the X- and Y-axis knob
tension
•
In the X-direction, hold the specimen holder
and move the stage guide to the left and right
until it hits the stoppers (Fig. 3-11/
1
,
2
).
•
In Y-direction, hold the upper stage and move it
to the front and rear until it hits the stoppers.
(2) Rotatable mechanical stage 75x50/240° R
NOTE
Rotatable stages can be used only with
the stage carrier for rotatable stages
(430710-9011-000).
a) How to remove the stage
•
Loosen the screw cap (
2
) of the spring
pin with approx. 3 turns.
•
Press the stage forward against the spring pin
(
1
), lift the back side off the stage
carrier (
3
) and remove it by lifting it
upward.
•
Tighten the screw cap (
2
).
Fig. 3-11
Adjusting the X- and Y-axis knob
tension
Fig. 3-12
Changing the rotatable mechanical
stage