Mounting/removing the 80x60
motorized mechanical stage on the
Axioscope 7 motorized material
stand
The motorized stage can only be mounted
on/removed from the stand as an entire unit with
the stage carrier (see section 3.1.15 on page 59).
How to remove the stage
•
Unplug the power cable connector (Fig. 3-20/
1
)
of the stage from the socket on the stand
(Fig. 3-20/
2
).
•
Turn the stage (Fig. 3-19/
1
) and stage carrier
6
) upside down and remove the four
mounting screws (Fig. 3-19/
5
) with the Allen
wrench (AF 3) (
4
).
•
Carefully remove the stage from the carrier.
How to mount the stage
Fig. 3-20
Plugging in the power cable of the motorized mechanical stage
•
Turn stage (
1
) and stage carrier (
6
) upside down carefully and match the threaded
holes in the bottom of the stage (Fig. 3-19/
2
) to the corresponding holes in the carrier (
3
).
•
Insert the four mounting screws (
5
) into the holes.
•
Align the stage in XY-direction and tighten the screws with the Allen wrench (AF 3).
•
Plug the power cable connectors (
1
) into the sockets on the stage (
3
) and on the
stand (
2
).
Fig. 3-19
Changing the motorized mechanical