DESCRIPTION OF THE INSTRUMENT
ZEISS
Controls and functional elements on microscope components Axioscope 5/7/Vario
40
430035-7344-001
03/2018
2.5.7
14x40 mm stop sliders for aperture
and field diaphragms
To adjust the reflected light beam path using the
KÖHLER method, one 14x40 mm stop slider
(
4
) mounted in the upper part of the
stand is required to function as a field diaphragm
(
F
) and another to function as the aperture
diaphragm (
A
).
•
To open the diaphragm, turn the knurled wheel
(
1
) up.
•
To close the diaphragm, turn the knurled wheel
(
1
) down.
•
To center the diaphragm to the beam path,
adjust the centering screws (
2
and
3
)
with an Allen wrench (AF 3).
NOTE
When applying fluorescent light, you may want to mount an FL attenuator (if not pre-
installed) on the level of the aperture diaphragm in order to attenuate the fluorescence
radiation.
Fig. 2-15
14x40 mm stop slider