MultiLab 4010-1W
Conductivity
ba76141e07
08/2018
61
8.3.6
Calibration data
The calibration record of the last calibration is available in the menu
Calibration
/
Calibration record
. To open it in the measured value display, press the
<CAL__>
key.
The calibration records of the last 10 calibrations are available in the menu
Cali-
bration
/
Calibration data storage
/
Display
. To open the
Calibration
menu in the
measured value display, press the
<ENTER>
key.
Calibration evalua-
tion
After calibration, the meter automatically evaluates the current status of the cal-
ibration. The evaluation appears on the display and in the calibration record.
The calibration record is automatically transmitted to the interface
after calibrating.
Menu item
Setting/
function
Explanation
Calibration
/
Calibration data
storage
/
Display
-
Displays the calibration records.
Further options:
Scroll through the calibration
records with
<
><
>
.
Output the displayed calibration
record to the interface with
<F2>
/
[USB output]
.
Output all calibration records to
the interface with
<F2__>
[USB
output]
.
Quit the display with
<F1>
/
[Back] or
<ENTER>
.
Switch directly to the measured
value display with
<M>
.
Calibration
/
Calibration data
storage
/
Output to
RS232/USB
-
Outputs the calibration data storage
to the interface (see section 12
T
Display
Calibration record
Cell constant [cm
-1
]
+++
Within the range
0.450 ... 0.500 cm
-1
Error
Error
Outside the range
0.450 ... 0.500 cm
-1
Error elimination (see section 14
W
Содержание MultiLab 4010-1W
Страница 105: ...MultiLab 4010 1W Index ba76141e07 08 2018 105 ...
Страница 110: ...Appendix MultiLab 4010 1W 110 ba76141e07 08 2018 ...
Страница 112: ...Contact Information MultiLab 4010 1W 112 ba76141e07 08 2018 ...
Страница 113: ......