MultiLab 4010-1W
Dissolved oxygen
ba76141e07
08/2018
51
Proceed as follows to calibrate the D.O. sensor:
1.
Connect the D.O. sensor to the meter.
2.
Immerse the D.O. sensor in the reference solution.
3.
Start the calibration with
<CAL>
.
4.
Start the measurement with
<ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
5.
Wait for the measurement with stability control to be finished (beep,
[HOLD][AR] status indicator).
6.
Using
<
> <
>
, set the measured value to adjust the displayed value
to the nominal value (value of the comparison measurement). Subse-
quently take over the adjustment with
<ENTER>
.
The meter switches to the measured value display.
The status display
[Factor]
is active.
O
2
7.92
mg/l
25.1
°C
Compare
03.04.2013
08:00
DO Adjust
O
2
7.92
mg/l
20.3°C
Compare
03.04.2013
08:00
DO Adjust