pH value
MultiLab 4010-1W
34
ba76141e07
08/2018
5.2.6
Calibration points
Calibration can be performed using one to five buffer solutions in any order
(single-point to five-point calibration). The meter determines the following
values and calculates the calibration line as follows:
5.2.7
Calibration data
22.
Set the nominal buffer value for the measured temperature with
<
><
>
.
23.
Accept the calibration value with
<ENTER>
.
The calibration display for the next buffer appears (voltage display).
24.
If necessary, use
<M>
to finish the calibration.
The calibration record is displayed.
or
Continue calibrating using the next buffer with
<ENTER>
.
After the fifth buffer has been measured the calibration is automat-
ically finished. Then the calibration record is displayed.
The calibration line is determined by linear regression.
Calibration
Determined
values
Displayed calibration data
1-point
Asymmetry
Zero point =
Asymmetry
Slope = Nernst slope
(-59.2 mV/pH at 25 °C)
2-point
Asymmetry
Slope
Zero point =
Asymmetry
Slope =
Slope
3-point to 5-
point
Asymmetry
Slope
Zero point =
Asymmetry
Slope =
Slope
The calibration line is calculated by
linear regression.
You can display the slope in the units, mV/pH or % (see section
9.1.1 S
The calibration record is automatically transmitted to the interface
after calibrating.