Settings
MultiLab 4010-1W
68
ba76141e07
08/2018
Calibration
/
Zero
calibration
(only for
4100 ProBOD, 5010
with 4011 Adapter-
ProOBOD)
Starts the 0-point calibration (see
section 7.3.6 Z
Calibration
/
Calibration interval
1 ...
180
...
999 d
Calibration interval
for the D.O.
sensor (in days).
The meter reminds you to calibrate
regularly by the flashing sensor
symbol in the measuring screen.
Calibration
/
Comparison meas.
on
off
Enables to adjust the measured
value with the aid of a comparison
measurement, e.g. Winkler titra-
tion.
For details, see section 7.3 C
Calibration
/
Sensor
cap coefficients
(only for ProOBOD)
K1
...
K5
KC
Enter the cap coefficients here after
exchanging the sensor cap.
For details, see section 9.3.2
E
Calibration
/
Membrane cap
(only for 4100 Pro-
BOD, 5010 with
4011 Adapter)
Black
Yellow
After exchanging the sensor cap,
select the cap type here.
FDO Check
/
Start FDO Check
(only for FDO 4410)
-
Starts the check with the
FDO Check
FDO Check
/
Check
interval
(only for FDO 4410)
1 ...
60
...
999 d
Interval for the
FDO Check
(in
days).
The meter reminds you to check
the sensor regularly by
FDO Check
status indicator in the measuring
screen.
Salinity
/
Sal correc-
tion
(only for param-
eter mg/l)
on
off
Manual salt content correction for
concentration measurements.
Salinity
/
Salinity
(only for parameter
mg/l)
0.0
... 70.0
Salinity or salinity equivalent for the
salt content correction.
This menu item is only available if
the manual salinity correction is
switched on.
Menu item
Possible
setting
Explanation