Dissolved oxygen
MultiLab 4010-1W
54
ba76141e07
08/2018
ibration. The evaluation appears on the display and in the calibration record.
Calibration evalua-
tion FDO 4410
Calibration evalua-
tion ProOBOD,
4100 ProBOD,
YSI 5010 with 4011
Adapter
Calibration record
(USB output,
example FDO 4410)
For evaluation, the characteristic curve of the sensor that was
determined is compared to the characteristic curve of an ideal sen-
sor at similar environmental conditions (relative slope S): S = S
sen-
sor
/ S
ideal sensor
. An ideal sensor has a slope of 1.
Display
Calibration record
Relative slope
+++
S = 0.94 ... 1.06
++
S = 0.92 ... 0.94
or
S = 1.06 ... 1.08
+
S = 0.90 ... 0.92
or
S = 1.08 ... 1.10
Error
Error
S < 0.90
or
S > 1.10
Error elimination (see section 14 W
Display
Calibration record
+++
Error elimination (see section 14 W
CALIBRATION Ox
03.04.2013 07:43:33
FDO 4410
Ser. no. 12B100016
SC-FDO
12B100015
Sensor
+++