10.3
Measurement data storage
10.3.1 Managing the measurement data storage
For each measurement data storage (automatic and manual) the following
functions are available:
Display
Output to RS232/USB
Erase
The storage is edited in the menu
Storage & config
/
Data storage
. To open the
Storage & config
menu, press the
<ENTER__>
key in the measured value
display.
Open the manual or automatic data storage directly with the
<RCL>
or
<RCL__>
key.
Settings
The settings are explained here using the manual data storage as
an example. The same settings and functions are available for the
automatic data storage.
Menu item
Setting/
function
Explanation
Data storage
/
Manual data storage
/
Display
-
Displays all measurement data-
sets page by page.
Further options:
Scroll through the datasets with
<
><
>
.
Output the displayed dataset to
the interface with
<F2>
/
[USB
output]
.
Quit the display with
<F1>
/
[Back].
Data storage
/
Manual data storage
/
Output to RS232/
USB
-
Outputs all stored measurement
data to the interface.
Data storage
/
Manual data storage
/
Erase
-
Erases the entire manual mea-
surement data storage.
Note:
All calibration data remain stored
when this action is performed.