pH value
MultiLab 4010-1W
40
ba76141e07
08/2018
sor evaluation.
Carrying out a QSC
control calibration
A QSC control calibration can, e.g. be useful if the sensor evaluation noticeably
changed (after some normal calibrations).
You can carry out QSC control calibrations at greater intervals than normal cali-
brations.
1 QSC scale
The double arrow on the QSC scale indicates the current sensor
evaluation
1
25.0
7.007
pH
°C
0
14
01.02.2014
08:00
i
USB output
1.
Prepare for calibration the meter, sensor and the buffer solutions of the
QSC-Kit.
2.
Open the menu for measurement settings with
<ENTER>
.
3.
In the
QSC
menu, select
Control calibration
with
<
><
>
.
The calibration display appears.
AutoCal QSC-Kit
is displayed as the
buffer.
Exclusively use the QSC-Kit for the QSC calibration. If you use other
buffers, you will have no valid QSC control calibration.
4.
Start the measurement with
<ENTER>
.
Calibration with the buffers of the QSC-Kit is done like a normal three-
point calibration (see section 5.2.4 C
As soon as the three-point calibration has been successfully carried out
you can decide whether to accept or discard the calibration as the QSC
control calibration.