pH value
MultiLab 4010-1W
36
ba76141e07
08/2018
Calibration record
(
USB output
)
-
-30 ... <-25
or
->+25 ... +30
-62.0 ... <-61.0
or
>-56.0 ... -50.0
Clean the IDS sensor according to the sensor operating manual
Error
Error
<-30
or
>+30
<-62.0
or
> -50,0
Error elimination (see section 14 W
For pH IDS sensors you can optionally enable a more finely graded
calibration evaluation (QSC) (see section 5.2.9 QSC
Display
Calibration
record
Zero point
[mV]
Slope [mV/pH]
4010-1W
Ser. no. 11292113
CALIBRATIONpH
01.02.2014 15:55
Ser. no. 10501234
YSI
Buffer 1
4.00
Buffer 2
7.00
Buffer 3
10.00
Voltage 1
184.0 mV
Voltage 2
3.0 mV
Voltage 3
-177.0 mV
Temperatur 1
24.0 °C
Temperatur 2
24.0 °C
Temperatur 3
24.0 °C
Slope
-60.2 mV/pH
Asymmetry
4.0 mV
Sensor
+++
etc...