pH value
MultiLab 4010-1W
30
ba76141e07
08/2018
Continuing with two-
point calibration
Continuing with
three- to five-point
calibration
9.
Thoroughly rinse the sensor with deionized water.
10.
Immerse the pH sensor in buffer solution 2.
11.
Start the measurement with
<ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
12.
Wait for the end of the measurement with stability control (beep,
[HOLD][AR] status indicator), or take over the calibrated value with
<ENTER>
.
The calibration display for the next buffer appears (voltage display).
13.
If necessary, finish the calibration procedure as a two-point calibration
with
<M>
.
The calibration record is displayed.
pH
0.0
24.8
°C
AutoCal YSI
7.000
01.07.2011
08:00
AR
Buffer
mV
14.
Thoroughly rinse the sensor with deionized water.
15.
Immerse the sensor in the next buffer solution.
16.
Start the measurement with
<ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
pH
180.0
24.8
°C
AutoCal YSI
4.000
01.07.2011
08:00
AR
Buffer
mV