control/alarm condition is processed by the 5X00 so long as alarms are disabled. If
alarms are enabled, the relay de-energizes when one of the four front panel softkeys
is pressed. Pressing one of four softkeys resets the
alarm system. The Alarm system becomes active again if, after the aux hold off
period, the alarm condition still exists.
Digital events (event log) are identified by aux/E-Aux number, input type (digital),
relay condition (on/off), and do not identify digital condition (open, or closed).
Aux digital events are not identified as control or alarm events regardless if alarms
are enabled.
E-Aux systems use an IOEM channel as the digital input source. See IOEM/REM
Safety and Installation guide for wiring and input channel configuration.
Configure Aux and E-Aux digital systems
Please review - Use AquaManager to configure system inputs and outputs - page 146.
1. Wire digital input at IOEM for E-Aux system - see IOEM/REM Safety and
Installation guide>Chapter 3>Step 5>Wire non temperature IOEM inputs;
OR
1. Wire Aux digital devices - page 49.
2. Wire relay output device to local instrument, or any networked 5X00
instrument - page 46.
OR
2. Wire relay output device to REM - see IOEM/REM Safety and Installation
guide>Chapter 3>Step 4>Wire REM Relay Outputs.
3. Configure system menu. For Aux - page 101. For E-Aux - page 107.
4. Enter Aux 1-2/E-Aux 1-8 in Sensor Setup menu
Menus → System → Sensor Setup → Aux 1-2
→ E-Aux 1-8
5. Use
▲
▼
to scroll and highlight submenus.
6. Press
to select.
E-AUX
E-AUX
Configuring the 5200A
7. Enable and configure submenus. See Aux/E-Aux Digital relay for digital
relay functionality - page 161. Configure relay using AquaManager - see
page 146.
8. Optional - Configure digital 4/20mA output - page 178.
9. Press
to save configuration and return to Run Screen.
10. Test system output device(s) at→System→Service→Output Test - page 141
or create an event to verify configured relay(s) energize.
Notes:
-
It is important to keep the 5X00 at the Run Screen during normal operation.
Alarm functionality (email alarms, activation of enabled alarm relays, alarm
icons, and buzzer) is suspended when 5X00 is not at Run Screen.
-
At sensor log Aux (1:1.00 ) is reporting open condition. Aux (1: 0.00) is re-
porting closed condition.
-
Do not enable aux digital alarms when the sensor is in the control/alarm con-
dition. Enabling aux digital alarms when sensor is in control/alarm condition
can result in aux system not properly processing existing alarm condition.
-
One relay can be assigned to multiple 5X00 systems. It is essential that the relay
configuration be verified in order to avoid conflict of operation of wired relay
devices. If not configured properly, assigning one relay to multiple systems
can result in invalid operation of 5X00.
-
Event logs are created when an aux relay energizes and de-energizes regardless
if alarms are enabled.
-
Aux alarms must be enabled to trigger General Alarm.
-
General Alarm may still be active even though aux system is no longer in alarm.
See aux relay information below and General Alarm - page 131.
-
E-Aux system datalogs are not accessible at the 5X00. They must be viewed
using AquaManager.
Aux and E-Aux Analog (0-1VDC, 0-5VDC and 4/20mA)
Menus → Sensor Setup→Aux 1-2
→E-Aux 1-8
Configure analog control and alarm low and high values, control and alarm relays
and hold off time in the Sensor Setup Menu.
E-Aux systems use an IOEM channel as the analog input source. See IOEM/REM
Safety and Installation guide for wiring and input channel configuration.
E-AUX
Configuring the 5200A