YSI 5200A
111
YSI 5200A
110
•
Use
▲
▼
to scroll and highlight IOEM channels.
•
Press
to select.
•
Press
to save settings and return to Run Screen.
Notes:
-
The Channel menu displays a generic list of instruments being Channel 1-8.
- The Channel menu will always display Channels 1-8 regardless if the
IOEM device selected is an IOEM-4 which has only 4 channels and not
8.
-
IOEM channel modes can only be configured using AquaManager. See
AquaManager online help for directions to configure IOEM channel input
mode (digital, 0-1V, 0-5V, 4mA-20mA, temperature);
Input Type
Menus → System → Configure E-Aux →E-Aux 1-8→Input Type →Digital
→0-1v
→0-5v
→4/20ma
→Temperature
Select the input type for the device(s) wired to IOEM channel(s). E-Aux Sensor Setup
menu selections are based on the E-Aux input type configured at system>E-Aux
menu. Input types are discussed in the following sections.
E-Aux Digital input type
Menus → System → Configure E-Aux →E-Aux 1-8→Input Type→Digital
Use the digital input when wiring an external device to control or alarm when the
contacts of the external switch are open or close. A common application would be
to detect high or low water levels using a float switch.
Configuring the 5200A
Notes:
-
IOEM input channel modes can only be configured using AquaManager. See
AquaManager online help for directions to configure IOEM channel input
mode (digital, 0-1V, 0-5V, 4mA-20mA, temperature);
-
Suffix, Min, Max, Input Scale Min, Input Scale Max and Display Format are
not available for digital inputs.
-
In the 5200A sensor log, aux digital “closed” is logged as Aux 0.00. Aux digital
“open” is logged as Aux 1.00.
Directions to configure E-Aux Digital - Configure E-Aux System menu:
1. Wire E-Aux digital devices - see IOEM Safety and Installation
Guide>Chapter 3 Installation and Wiring>IOEM input configuration>Wire
digital input.
2. Refer to and follow directions at IOEM Safety and Installation
Guide>Chapter 4>Quick Start - IOEM Configuration to confirm correct
network connectivity;
3. Follow directions in the Quick Start section to configure IOEM channel
mode to digital.
4. Enter E-Aux menu
5.
Menus → System → Configure E-Aux →E-Aux 1-8
6. Use
▲
and
▼
to scroll and select submenus.
7. Press
to select.
8. Configure E-aux system name - page 108, instrument - page 108, chan-
nel - page 109 and input type - page 110.
9. Press
to save configuration and return to Run Screen.
10. Go to Sensor Setup Menus → Sensor Setup E-Aux 1-8 - page 162 to en-
able and configure control/alarm on condition, control/alarm relay, and
hold off time.
11. Press
to save configuration.
Configuring the 5200A