YSI 5200A
117
YSI 5200A
116
Data Logging
Menus → System →Data Logging
The 5200A saves information in the calibration, sensor, and event logs. Press
at the Run Screen to view the 25 most recent records for each log. The 5200A sen-
sor log stores 2000 records. The calibration log can store 500 records. The event
log can store 1250 records. Once a 5200A log is full, it will begin overwriting the
oldest records with new records (data wrapping).
Notes:
-
Use AquaManager to store and analyze 5200A logs. To avoid gaps in data use
polling intervals based on the user defined data logging configuration. Set the
data log interval so that analyzing the data is meaningful.
-
Aqu a Man age r u s e rs c an v i e w C on f i g u r at i on e ve nt s i n t h e
Configuration log. Configuration
events are created when certain 5200A
configuration changes are made. The configuration log is not accessible at the
5200A. See AquaManager Online Help for additional information.
Data Logging
Menus → System →Data Logging→Data Logging
The 5200A sensor log stores records for sensors and aux systems. Sensor logs are
time and date stamped based on the user configured logging interval. The logging
interval is user configured at (10-255 seconds or 1-255 minutes or hours).
Configuring the 5200A
In the above sensor log screen shot ORP is disabled.
1. Enter Data Logging menu
Menus → System →Data Logging→Data Logging
2. Use
▲
and
▼
to scroll and highlight submenus.
3. Press
to select.
4. Enable and configure submenus.
5. Press
to save configuration and return to Run Screen.
Press
to view the 25 most recent sensor logs.
Notes:
-
Temperature values are stored in °C regardless of display format.
-
Aux digital “closed” is logged as Aux 0.00. Aux digital “open” is logged as
Aux 1.00.
-
The 5200A sensor log records 0.00 resolution for aux analog inputs regardless
of the selected display format.
-
DO values are stored in mg/L & %Sat regardless of selected display format.
-
Sensor logs record the default system name and not the user configured
system name.
-
Changes to the data logging interval take effect after the expiration of the
interval being changed. Example - if the interval is changed from one hour
to one minute, one minute logging will begin after the expiration of the cur-
rent one hour interval. Power cycle the 5200A to make a change to the data
logging interval occur immediately.
-
Menu configurations are saved to the system by pressing
to return
to the Run Screen and not when the change is made in the menu. Excep-
tion - Date and time and clear data logs are updated immediately to the
system, not when
is pressed
.
-
The sensor log contains values for all sensor and aux systems regardless if the
system is disabled.
-
5200A sensor log values for disabled systems and systems without sensors
are not valid.
-
AquaManager data log displays 5200A conductivity values in µS to 0.00
resolution.
-
E-Aux system datalogs are not visible at the instrument. E-Aux system datalogs
are only viewable using AquaManager.