Feed % (of) Biomass
Menus →Timers→Feed Smart Timer→Feeder→FCR→Feed % Biomass
Enter the percentage of the total livestock weight (biomass) that will be used to
determine the daily feed weight. The value is user selectable 0.0–99.9%.
Daily (Feed) Weight
Menus →Timers→Feed Smart Timer→Feeder→FCR→Daily Weight
Configure the initial daily feed weight by multiplying the biomass amount by the
Feed % Biomass percentage (the 5200A will not calculate the initial amount). Ex-
ample—if the biomass is 1000 and the feed % biomass is 3%, then the initial daily
feed weight amount is 30. On each following day, the 5200A will use the biomass
value along with the feed % biomass to calculate the new daily feed weight amount.
Note:
-
The initial daily weight value must be calculated as described above. Config-
uring an invalid initial daily weight will result in incorrect biomass and daily
weight calculations when the 5200A crosses midnight.
FCR Factor
Menus →Timers→Feed Smart Timer→Feeder→FCR→FCR Factor
Enter the ratio of the amount of food required versus the corresponding increase
in fish weight (biomass). This value is used to automatically increase the biomass
amount—as shown in the example below.
Sensors
Menus →Timers→Feed Smart Timer→Feeder→Sensors
→Disabled/Enabled
→DO Low/High/DO % Change
→pH Low/High/pH % Change
→Temp Low/High/Temp % Change
Changes in water conditions affect the amount of food fish eat which in turn affects
fish growth. With the sensor feature enabled, the feed timer system will verify that
DO, temperature, and pH values are within user configured ranges. This verification
is done immediately prior to starting each feed duration in a feed cycle. If values
are reported that fall outside the acceptable configured ranges, the feed timer relay
energizes for a shorter period (user configured percentage) thereby reducing the
amount of food that is dispensed. Feed timer sensor ranges are independent of
sensor setup configuration ranges. For timed feed cycles, the feed interval for the
entire feed cycle is determined by duration of the first feed.
Example - The feed timer is configured as follows: sensor temperature low value
is 25.1oC, feed timer temperature change is 9%, timed feed mode, six feeds, daily
weight 2 units start time 8A, end time 9A, six 9.999 second feeds of .3333 units
each to occur at 9 minute 50 second intervals without sensor reduction. Prior
to 8A feed cycle, the start temperature value is reading 24.9oC. Feed duration of
9.999 seconds will be reduced by approximately 1 second (9%) because of the “out
of range” temperature value.
DO Low/pH Low/Temp Low
Enter the lower limit of the sensor range where no feed reduction will take place.
DO High/pH High/Temp High
Enter the high limit of the sensor range where no feed reduction will take place.
Configuring the 5200A
Configuring the 5200A