Ethernet
Menus → System → Communications → Ethernet
The serial to ethernet device speed and format are self configured at power up (TCP/
IP connections). If the serial to ethernet device is not installed or is not installed
properly, the ethernet menu will display (Not Available).
Notes:
-
5200A email alarming from a serial to ethernet device requires access to an
SMTP server accessible by the local network.
-
Do not reconfigure internal serial to ethernet device settings.
1. Install serial to ethernet device according to instruction provided by YSI
with the module.
2. The Ethernet icon
displays at the Run Screen when the serial to ethernet
device is properly installed.
3. Enter Ethernet menu:
Menus → System → Communications → Ethernet
4. Use
▲
and
▼
to scroll and highlight submenus.
5. Press
to select submenu.
6. Configure submenus as needed.
7. Press
to save configuration and return to Run Screen.
Notes:
-
The CC/Email address must be configured when using 5200A email alarming.
Specific alarm system information will not be sent if the CC/Email address
menu is not configured with a valid email address.
-
If Autopoller is being used for AquaManager email alarming on a TCP/IP
connection, it is strongly recommended that the 5200A email alarming using
the serial to ethernet device be disabled. When the ethernet device sends
alarm emails, all Autopoller SMS communication is suspended. This could
potentially delay any AquaManager alarm emails from being sent.
Configuring the 5200A
-
To disable the serial to ethernet device from sending alarms, clear all four
email addresses from the Communication → Ethernet
menus. Clear the
email addresses by using the “Spc” (space) key on the numeric keypad. Save
change at numeric keypad and press
to save configuration and return
to Run Screen.
DHCP
Enable DHCP if you are using the serial to ethernet device only for 5200A email
alarming and not for AquaManager remote TCP/IP access.
Notes:
-
When DHCP is enabled, the 5200A IP address is not updated at the IP ad-
dress menu.
-
When DHCP is enabled, the serial to ethernet device acquires SMTP and DNS
addresses during initialization.
-
AquaManager requires a static IP address; therefore, DHCP must be disabled
when using AquaManager. Consult the network administrator to obtain a
static IP address.
-
Power cycle the 5200A whenever DHCP is enabled.
IP Address
Menus → System → Communications → Ethernet → IP Address
The IP address can only be configured when DHCP is disabled.
Notes:
-
AquaManager users should program the Gateway address either by force se-
rial port in the AquaManager Communication menu or locally at the 5200A.
Poll node using RS232 communication to upload configuration information
including: static IP, port, subnet, and gateway addresses.
-
When DHCP is enabled, the IP address is not updated at the IP address menu.
-
Power cycle the 5200A whenever a new IP address is configured.
-
Numbers between 0-255 are valid for each of the four (4) IP Address fields.
Configuring the 5200A