YSI 5200A
147
YSI 5200A
146
Notes:
-
Always manually poll the 5200A after uploading configuration changes and
verify the uploaded information is retained. When configuration changes are
uploaded using AquaManager at the same time configuration changes are being
made by a user in the menus on the 5200A, the AquaManager uploads may
not be saved depending on the timing of keystrokes at the 5200A.
-
The Run Screen must be displayed during normal operation.
If the 5200A
is not at the run screen during AquaManager uploading, pressing
at
the 5200A will only save certain configuration changes.
See Chapter 5 AquaManager - page 198 and AquaManager online help for ad-
ditional information.
Sensor Setup
Menus → Sensor Setup →Aux 1 - 2
→E-Aux 1 - 8
→Conductivity
→DO
→ORP
→pH
→Temperature
→Salinity
Use AquaManager to configure system inputs and outputs
Any networked 5X00 and REM instrument relay can be configured to any networked
5X00 sensor control or alarm system. In addition, any E-Aux system (see page 107)
can be configured to use any networked IOEM input channel as its input source.
See local versus remote instruments on a network - page 86. Because of differ-
ences in how system relay outputs and E-Aux system inputs are configured locally
at a 5X00 instrument versus using AquaManager,
it is strongly recommended that
AquaManager be used to configure all 5X00 sensor setup menus. AquaManager
contains a number of software checks that help to validate E-Aux input and 5X00
system output relay assignment configuration. These same checks do not exist
when configuring systems directly at a 5X00 instrument.
AquaManager maintains
an up to date table of all node network instruments. This table gives AquaManager
the ability to display only networked instruments at E-Aux source input and system
relay output pulldown menus. (This AquaManager table is not visible to the user.)
An Example of configuring a 5X00 system control relay at an instrument versus using
AquaManager is provided below. The node network used in the example includes
a master (being a 5500D-04) and 4 slaves (slave 2 being a 5500D-02; slave 4 being
a 5400; slave 11 being a 5200A; slave 16 being an REM).
NEW
Configuring the 5200A
Instrument configuration
- the control relay instrument menus displays a generic
list of instruments. See instrument and relay name conventions - page 133. Any
instrument from the generic list can be selected/configured. This generic list in no
way accurately depicts the instruments on the actual node network. In this example,
the node network consists of a master and slave 2, slave 4 and slave 11 and slave 16,
yet at the system control instrument menu a generic list of instruments is displayed.
Which generic list includes Local Machine, Master and Slaves 1-31 and not just the
4 instruments on the node network. The relay list also displays a generic list which
displays relays 1-8 for all instruments. This generic list list does not take into account
that 5200As and REMs have 4 relays and not 8 relays. instruments. See screen shots
below of instrument and relay menus at instrument.
Control system instrument menu Control system relay menu
AquaManager configuration
- in the alternative, AquaManager knows what in-
struments are actually on the node network and consequently knows actual relays
available on each instrument. In this example, only the Master 5500D-04 and slaves
on the network are available at the instrument menu pull down. Each instrument
also will display only available relays based on the instrument type. In the relay menu
screen shot below 4 relays are displayed because Slave 16, an REM, was selected. See
screen shots below for AquaManager pull downs.
Control system instrument menu Control system relay menu
Configuring the 5200A