Directions for Use
48
PANOURA 18S Operation Manual
Radiation status
Function
To which mode is switched when Select
switch is pressed
AUTO
Specifies automatically the values for Tube
voltage and Radiation time appropriate for the
patient's tooth row size.
Manual
Auto
MANUAL
Allows the operator to modify arbitrarily the
value for Tube voltage, Tube current, or
Radiation time.
3
In the Manual status, press the Fix
switch, select “Tube voltage”, “Tube
current”, and “acquisition time”, and use
the “+” and “-” switches to modify the
settings.
To which mode is switched when Confirm
switch is pressed
Shifting to new setting
for tube current
Shifting to new setting
for tube voltage
Settings confirmed
Shifting to new setting
for irradiation time
See also
: "Parameter settings table"
P.26
4
In the main screen of Image Creator,
set up the tooth alignment size and
tooth alignment shape according to the
patient. Set the tomographic position at
0.
5
Click “New” in the main screen of Image
Creator and Image Acquisition Condi-
tion Setting screen is displayed.
Make sure that the settings in the
screen match with the settings made on
the Control Panel, and press “OK”.
Содержание PANOURA 18S
Страница 2: ......
Страница 23: ...Before Using this Product 21 PANOURA 18S Operation Manual...
Страница 24: ...Before Using this Product 22 PANOURA 18S Operation Manual...
Страница 44: ...MEMO...
Страница 108: ...MEMO...
Страница 118: ...MEMO...
Страница 120: ...MEMO...
Страница 122: ...MEMO...
Страница 124: ...MEMO...
Страница 129: ...Main specification 127 PANOURA 18S Operation Manual 11 3 2 Standing position wall mount short type Panorama type Unit mm...
Страница 130: ...Main specification 128 PANOURA 18S Operation Manual 11 3 3 Standing position wall mount long type Panorama type Unit mm...
Страница 137: ...Main specification 135 PANOURA 18S Operation Manual 11 3 10 Standing position wall mount Cephalo attached type Unit mm...
Страница 150: ...MEMO...
Страница 154: ...MEMO...
Страница 155: ......