Qualification Assurance (QA) Procedure
150
PANOURA 18S Operation Manual
12.2 QA Procedure
NOTE
: Perform an inspection according to following steps on each half of year and save
the result.
1
Attach the QA phantom to the chin rest block base.
phantom
Chin rest base
2
Acquire image of the QA phantom under the following conditions.
Image acquisition mode setting
Image acquisition mode
Standard panorama
Image acquisition status
Manual
Tube voltage
58kV
Tube current
2mA
Radiation time
16sec
User program mode setting
Cervical spine correction
OFF
Image Creator setting
Tooth alignment size
Normal
Tooth alignment shape
Normal
Tomographic position
0mm
Cervical spine correction
OFF
NOTE
: For the above image acquisitions, do not move the arm position. If you have
moved the arm position using the forward/back switch, press the reset switch to
return the arm to the original position before acquiring images.
Содержание PANOURA 18S
Страница 2: ......
Страница 23: ...Before Using this Product 21 PANOURA 18S Operation Manual...
Страница 24: ...Before Using this Product 22 PANOURA 18S Operation Manual...
Страница 44: ...MEMO...
Страница 108: ...MEMO...
Страница 118: ...MEMO...
Страница 120: ...MEMO...
Страница 122: ...MEMO...
Страница 124: ...MEMO...
Страница 129: ...Main specification 127 PANOURA 18S Operation Manual 11 3 2 Standing position wall mount short type Panorama type Unit mm...
Страница 130: ...Main specification 128 PANOURA 18S Operation Manual 11 3 3 Standing position wall mount long type Panorama type Unit mm...
Страница 137: ...Main specification 135 PANOURA 18S Operation Manual 11 3 10 Standing position wall mount Cephalo attached type Unit mm...
Страница 150: ...MEMO...
Страница 154: ...MEMO...
Страница 155: ......