![background image](http://html1.mh-extra.com/html/yoshida/panoura-18s/panoura-18s_operation-manual_3502917058.webp)
Directions for Use
56
PANOURA 18S Operation Manual
Suppl.
: Wrong median beam alignment may affect the images as described below.
1. Inclined patient’s face
Problems
• Prominent difference between the tops of
left and right temporomandibular joints.
• Distorted image with a difference in size
between left and right mandibular angles.
Possible causes
• Inclined patient’s face toward either of
the left and right sides.
Corrective measure
• Align the reference median line (vertical
line) to the patient’s midsagittal line.
2. One half of the patient’ face enlarged and the other reduced
Problems
• One half of the patient’s face is enlarged
and the other reduced.
• A difference in size between the tops of
left and right temporomandibular joints.
Possible causes
• The patient does not face front, deflecting
toward either of the left and right sides.
• Inclined patient’s face toward either of
the left and right sides.
Corrective measure
• Align the reference median line (vertical
line) to the patient's midsagittal line.
Содержание PANOURA 18S
Страница 2: ......
Страница 23: ...Before Using this Product 21 PANOURA 18S Operation Manual...
Страница 24: ...Before Using this Product 22 PANOURA 18S Operation Manual...
Страница 44: ...MEMO...
Страница 108: ...MEMO...
Страница 118: ...MEMO...
Страница 120: ...MEMO...
Страница 122: ...MEMO...
Страница 124: ...MEMO...
Страница 129: ...Main specification 127 PANOURA 18S Operation Manual 11 3 2 Standing position wall mount short type Panorama type Unit mm...
Страница 130: ...Main specification 128 PANOURA 18S Operation Manual 11 3 3 Standing position wall mount long type Panorama type Unit mm...
Страница 137: ...Main specification 135 PANOURA 18S Operation Manual 11 3 10 Standing position wall mount Cephalo attached type Unit mm...
Страница 150: ...MEMO...
Страница 154: ...MEMO...
Страница 155: ......