Directions for Use
45
PANOURA 18S Operation Manual
2.3 Procedure for X-ray image acquisition
2.3.1 Turning the power switch on
1
Turn the computer’s power switch on.
2
Turn the power switch on (a | symbol will appear) on the equipment to check that the
power LED (green) has gone on (also, check that the Emergency Stop switch has been
released).
At power-on, the arm unit will move automatically to the patient introduction point (The set-
tings for automatic motion may be modified in the program mode. See “Table 4 Descriptions
of user program mode” of the section “4 User program mode”).
Caution
• At power-on, usually the arm unit will move automatically to the patient introduction
point. Don’t put anything within the movable range of the arm unit, otherwise its sur-
rounding physical proprietaries may be damaged.
• Ask the patient to keep off the equipment until the arm unit returns to its reset point.
• If the equipment is not used for a long time, turn off the power switch (a O symbol will
appear)
- Properly ground the computer.
- Before image acquisition, make sure that the sensor (detachable) is installed and
locked at the docking station (panoramic).
3
Wait for a while until the link between the internal sensor and the computer has been
established.
NOTE
: If the link between the sensor and the computer is not established for more
than one minute, turn off the equipment and turn it on once again.
Содержание PANOURA 18S
Страница 2: ......
Страница 23: ...Before Using this Product 21 PANOURA 18S Operation Manual...
Страница 24: ...Before Using this Product 22 PANOURA 18S Operation Manual...
Страница 44: ...MEMO...
Страница 108: ...MEMO...
Страница 118: ...MEMO...
Страница 120: ...MEMO...
Страница 122: ...MEMO...
Страница 124: ...MEMO...
Страница 129: ...Main specification 127 PANOURA 18S Operation Manual 11 3 2 Standing position wall mount short type Panorama type Unit mm...
Страница 130: ...Main specification 128 PANOURA 18S Operation Manual 11 3 3 Standing position wall mount long type Panorama type Unit mm...
Страница 137: ...Main specification 135 PANOURA 18S Operation Manual 11 3 10 Standing position wall mount Cephalo attached type Unit mm...
Страница 150: ...MEMO...
Страница 154: ...MEMO...
Страница 155: ......