Before Using this Product
26
PANOURA 18S Operation Manual
Name
Function
4
"+" and "-" switch
"Used to specify any value for "Tube voltage", "Tube current", and ""Radiation
time"". ""+"" increases the current value by one level and "-" decreases by one
level.
For more information on the allowable ranges of tube voltage, tube current,
and radiation time, see "Parameter settings table" P.26 ."
5
Reset switch
Rotates the arm up to the start point for each radiation mode. Also used to
release the error message.
Pressing this switch for 3 seconds rotates the arm to the sensor detachable
position.
6
Positioning beam switch
Positions the patient by means of positioning beams. When pressed, moves
the arm unit to the positioning point and turn on the positioning beams associ-
ated with "Meadina line", "oculoauricular", and "Front tooth topographic area".
The beams will go off 60 seconds after automatically. When pressed again
during the beam on, the beam will go off.
7
Up switch
When pressed and held down, moves the sliding mechanism up to adjust the
chin rest level. When released, stops sliding movement.
8
Down switch
When pressed and held down, moves the sliding mechanism down to adjust
the chin rest level. When released, stops sliding movement.
9
Forward switch
When pressed and held down, moves the arm unit forward (toward the
column) to adjust the patient position for acquiring topographic images of front
teeth. When released, stops sliding movement.
10 Back switch
When pressed and held down, moves the arm unit backward (toward the
opposite side of the column) to adjust the patient position for acquiring topo-
graphic images of front teeth. When released, stops sliding movement.
* Pressing the positioning beam switch causes the positioning beam to remain activated for 60 seconds.
Parameter settings table
Parameter
Settings
Tube voltage [kV]
58
60
63
65
68
71
73
76
79
82
Tube current [mA]
2.0
2.5
3.2
4.0
5.0
6.4
8.0
10
Radiation time
[sec]
Panoramic
Adult
8
14
16
Child
6.4
11.2
12.8
TMJ
8(*1)
Single
0.10
0.12
0.16
0.20
0.25
0.32
0.40
0.50
0.63
0.80
1.00
1.20
1.60
2.00
2.50
3.20
Cephalo* PA, Carpus
8.0
Carpus*
LA
10.0
*For a Cephalo-attached equipment
*Radiation time for TMJ image acquisition is 4 seconds x 2. When acquiring images of both mouth opening and
closing positions, it will be two acquisitions of 4 seconds x 2 each.
Содержание PANOURA 18S
Страница 2: ......
Страница 23: ...Before Using this Product 21 PANOURA 18S Operation Manual...
Страница 24: ...Before Using this Product 22 PANOURA 18S Operation Manual...
Страница 44: ...MEMO...
Страница 108: ...MEMO...
Страница 118: ...MEMO...
Страница 120: ...MEMO...
Страница 122: ...MEMO...
Страница 124: ...MEMO...
Страница 129: ...Main specification 127 PANOURA 18S Operation Manual 11 3 2 Standing position wall mount short type Panorama type Unit mm...
Страница 130: ...Main specification 128 PANOURA 18S Operation Manual 11 3 3 Standing position wall mount long type Panorama type Unit mm...
Страница 137: ...Main specification 135 PANOURA 18S Operation Manual 11 3 10 Standing position wall mount Cephalo attached type Unit mm...
Страница 150: ...MEMO...
Страница 154: ...MEMO...
Страница 155: ......