Qualification Assurance (QA) Procedure
151
PANOURA 18S Operation Manual
3
Check the following from the acquired panoramic images.
1. Check the circularity of the steel ball in the center.
Radius of the X-direction should be within ±19% of the radius of Y-direction.
X-direction
Y-direction
See also
: For measurement functions, see the viewer manual.
2. Check the distance from the steel ball in the center to the steel balls in the right/left
end.
For the measurement values A and B below, it should be |(A-B)/(A+B)|
≦
0.035.
A
B
See also
: For measurement functions, see the viewer manual.